Is computer-assisted total knee replacement for beginners or experts? Prospective study among three groups of patients treated by surgeons with different levels of experience

Materials and methods

Three homogeneous groups who underwent computer-assisted TKR were included in the study: group A [surgery performed by a surgeon experienced in both TKR and computer-assisted surgery (CAS)], B [surgery performed by a surgeon experienced in TKR but not CAS], and C [surgery performed by a general orthopedic surgeon]. In other words, all of the surgeons had different levels of experience in TKR and CAS, and each group was treated by only one of the surgeons. Cutting errors, number of re-cuts, complications, and mean surgical times were recorded. Frontal femoral component angle, frontal tibial component angle, hip–knee–ankle angle, and component slopes were evaluated.

Results

The number of cutting errors varied significantly: the lowest number was recorded for TKR performed by the surgeon with experience in CAS. Superior results were achieved in relation to final mechanical axis alignment by the surgeon experienced in CAS compared to the other surgeons. However, the total number of outliers showed no statistically significant difference among the three surgeons. After 11 cases, there were no differences in the number of re-cuts between groups A and C, and after 9 cases there were no differences in surgical time between groups A and B.

Conclusion

A beginner can reproduce the results of an expert TKR surgeon by means of navigation (i.e., CAS) after a learning curve of 16 cases; this represents the break-even point after which no statistically significant difference is observed between the expert surgeon and the beginner utilizing CAS.

Introduction

Nowadays, to obtain the best positioning of the components during a total knee replacement (TKR), the surgeon must use the best technology. Malalignment can adversely effect the longevity of knee prostheses, causing early wear and implant loosening, both of which are linked to suboptimal implant positioning [1–3]. Greater than 3° of varus or valgus malalignment in total knee replacement can result in higher failure rates, whilst correct alignment has been associated with improved clinical outcome [3–5]. Several authors have shown that traditional hand-guided alignment systems can produce potential errors in the bone-cutting process, even when used by experienced surgeons [6–11]. The use of a navigation system could help the young surgeon and the expert surgeon to achieve good, long-lasting results.

Recently, Manley et al. [12] showed that patients undergoing TKR in low-volume hospitals (1–25 procedures/year) had a higher risk of early revision at five and eight years compared with those performed in hospitals with the highest volumes (>200 procedures/year). Total knee replacement performed with computer-aided alignment appears to produce superior radiological results to hand-guided techniques [13–15]. These computer-assisted surgery (CAS) systems have been shown to both improve mechanical alignment and reduce outliers. Both of these outcomes are linked to a potential decrease in the TKR revision rate. Computer navigation provides continuous feedback during all phases of knee replacement surgery, providing an opportunity to correct any bone-cutting errors. Using a navigation system implies the application of an authentic protocol. There are obligatory steps to be carried out when using the computer; every surgeon must perform the same steps, and every step is recorded, so we can check what has been performed with precision. It has been suggested that computer navigation could be used as a teaching tool, so that even an inexperienced TKR surgeon would be able to perform more expertly [6, 16, 17]. In 2008, in a retrospective study, Yau et al. failed to show any improvement in postoperative alignment using a computer-assisted technique in a low-volume knee practice [18]. In 2005, Daubresse et al. hypothesized that the learning curve for a computer-navigated TKR technique cannot be longer than that of the free-hand technique, even in a community hospital [19].

The aim of this study was to determine the break-even point for different experiences in TKR and CAS. The frequency of intraoperative bone cut errors, the final implant alignment, and the surgical time were assessed to evaluate the learning curve.

Materials and methods

A prospective study of 75 selected patients undergoing computer-assisted TKR was undertaken. Strict inclusion criteria were adopted in the study; only patients with primary osteoarthritis, a body mass index of ≤35, a maximum mechanical axis deformity of less than 15°, and at least 90° of knee flexion were included. Before the study we assigned each patient into one of three equally sized groups (groups A, B, and C; see Table 1). Each patient was informed and gave consent prior to being included in the study. The study was authorized by the local ethical committee and was performed in accordance with the ethical standards of the 1964 Declaration of Helsinki as revised in 2000.

Group A had their surgery performed by a surgeon experienced in both knee replacement (more than 70 TKRs/year) and computer-navigated (more than 250 CAS-TKRs implanted) surgery. The surgeon for group B patients was experienced in knee replacement (more than 70 TKRs/year) but had not previously performed computer-guided surgery. A general orthopedic surgeon performed all TKRs in group C. This surgeon had a low-volume TKR experience (less than 20 TKRs/year) and no previous experience with computer-guided surgery. The same computer navigation system (Vector Vision, version 1.52, BrainLAB, Munich, Germany) was used in all TKRs. The prosthesis used in all patients was the same (Genesis II, Smith and Nephew, Memphis, TN, USA). Each surgeon involved in the trial was never involved in the operations of the other two.

A standardized operative approach was followed in all TKRs. In all cases, the midline patellar skin incision was pre-drawn to a length ranging from between 13.5 and 15.5 cm. A medial para-patellar arthrotomy was extended proximally to the quadriceps tendon in all patients. The patellar was retracted laterally in each case. All prostheses were implanted using the same dedicated instruments, including cutting blocks specifically designed for computer navigation. The cutting blocks were fixed with a combination of treated and smooth pins in all patients. All implants were cemented. No patients underwent patellar resurfacing. The same pre- and postoperative rehabilitation protocols were used in all three groups. Early weight-bearing was encouraged in all patients if tolerated.

For each implant, the tibial and femoral cutting errors and the number of re-cuts were recorded. A cutting error occurred when the planned angle of the bone cut as measured by the cutting block differed from the angle seen after sawing. The cutting error was measured in the frontal and sagittal planes of both the tibia and femur, giving four measurements (Fig. 1). According to a pre-determined surgical protocol, a re-cut was required if the cutting error was ≥3°. The number of complications and the mean surgical time (time between skin incision and tourniquet release) were measured for each group.

Standing radiographs were performed six months postoperatively with the knee in maximum extension, the patella pointing forward, and both hips and ankles visible on the film. The lateral radiographs were taken with the knee in 30° of flexion on a radiographic film. The radiographs were taken according to a standardized protocol and magnification. If malrotation was detected the radiographs were repeated. An independent radiologist assessed all radiographs.

The frontal femoral component (fFC) angle, the frontal tibial component (fTC) angle, the hip–knee–ankle (HKA) angle, and the sagittal orientations of components (slopes) were all measured. These parameters were utilized to evaluate the quality of the surgical outcome. The fFC is the angle between the mechanical axis of the femur and the transverse axis of the femoral component. The fTC is the angle between the mechanical axis of the tibia and the transverse axis of the tibial component. The slopes of the femoral (FS) and tibial (TS) components were defined as the angle between a line drawn tangential to the base plate (surface in contact with bone) of the respective component and the anterior femoral cortex or tibial mechanical axis, respectively (Figs. 2, 3). The desired prosthesis alignment for each parameter was determined prior to the study as an fFC angle of 90°, fTC angle of 90°, HKA angle of 180°, femoral slope of 90°, and tibial slope of 87°. The total number of outliers for all five radiological parameters were determined for each group and compared. An outlier was defined as a postoperative malalignment of any parameter of greater than 3° from the target value (Table 2).

Statistical analysis of the results was performed and the three groups were compared. Because of an abnormal data distribution, nonparametric testing (Kruskal–Wallis test) was performed using Statistica 7.0 software (StatSoft Inc., Tulsa, OK, USA) for analysis. Statistical significance was set at p ≤ 0.05.



Results

Analysis of the demographic data for all three groups showed no statistically significant differences in preoperative flexion, body mass index, or preoperative deformity.

There were no complications relating to the surgical technique or the surgeon’s experience.

Statistically significant differences were, however, seen when the following parameters were compared among groups: distal femoral cut, proximal femoral cut, femoral component slope, mechanical axis. Statistically significantly inferior results were seen for the patients operated on by the general orthopedic surgeon concerning the distal femoral cut in the sagittal plane, compared to the other two groups (p = 0.05). No significant difference was seen for the distal femoral cut in the coronal plane among groups A, B, and C. For the proximal tibial cut in the coronal plane, standard deviations of 0.91°, 1.31°, and 1.28° were noted for groups A, B, and C, respectively. These differences were not statistically significant. A statistically significant difference (p = 0.007) was seen in the proximal tibial cut in the sagittal plane between the patients operated on by the surgeon experienced in computer-guided and knee replacement surgery cost in indore and the general orthopedic surgeon  in indore.

A statistically significant difference (p = 0.05) was seen in the femoral component slope between the patients operated on by the experienced TKR surgeons and the general orthopedic surgeon. The slope of the femoral component was 90.36° (range 87–94°), 89.92° (range 88–95°), and 90.68° (range 88–94°) in groups A, B, and C, respectively. There was no significant statistical difference in the postoperative fFC and fTC angles across the three patient groups.

The slope of the tibial component in group A was 86.72° (range 84–91°), in group B it was 87.44° (range 84–92°), and in group C it was 88.24° (range 84–91°). A statistically significant difference (p = 0.007) was noted between groups A and C. The patients who underwent TKR performed by the surgeon experienced in both computer-guided and knee replacement surgery had a statistically significantly improved mechanical axis when compared to the patients from groups B (p = 0.030) and C (p = 0.0006) (Table 3; Fig. 4). Despite these findings, no statistically significant difference was seen among the three patient groups in terms of the total number of outliers for all five radiographic parameters.

There was a correlation between the level of experience in both computer navigation and knee replacement surgery and the number of re-cuts. Four re-cuts were seen in group A, eight re-cuts were needed in group B, and 13 re-cuts were done in group C. A statistically significant difference was seen between groups A and C (p = 0.02). This difference suggested an inverse relationship between the surgeon’s experience and the number of re-cuts. We found no statistical difference between the group operated on by the CAS-trained surgeon and the group operated on by the TKR-trained surgeon, and the break-even point between the group operated on by the CAS-trained surgeon and the group operated on by the general orthopedic surgeon corresponded to 11 cases (Fig. 5).

A statistically significant increase in surgical time was seen for the patients in groups B and C (who had TKR performed by surgeons lacking experience in computer-assisted techniques) compared to group A. We observed no statistical difference among the surgeons after nine cases between group A and group B, and after 16 cases between group A and group C.


Summarizing, in group A, we observed statistically significantly superior results regarding the distal femoral cut, the proximal tibial cut, the mechanical axis, the number of re-cuts, and the surgical time when compared with group C; we also noted statistically significantly superior results concerning the mechanical axis and surgical time for group A compared to group B. We saw a statistically significantly improved mechanical axis in group B compared to group C.

No complications were seen in any of the three groups.

Discussion

Malalignment of a TKR has been shown to adversely influence implant survival. Different intraoperative pitfalls can affect the final postoperative alignment in TKR. Malalignment in the sagittal plane in excess of 3° can increase the implant failure rate and result in poorer clinical outcomes [3, 12]. Using traditional intramedullary alignment systems, deviations of up to 8° can occur in the femoral axis, depending on the size and length of the intramedullary guide [20]. In 2001, Mahaluxmivala et al. showed that TKR alignment improves with surgical experience [8]. Unstable cutting blocks and saw deviations during osteotomy have been shown to result in cutting errors [6, 10]. A strict correlation has been demonstrated between surgical experience of TKR and implant survival [3, 8, 12].

Computer-assisted surgery provides the surgeon with continuous intraoperative feedback on cutting errors and implant alignment during all phases of TKR [6]. Recent studies have demonstrated that computer navigation may play a role in reducing the learning curve in joint replacement surgery [17, 21].

The aim of the current prospective controlled trial was to assess the influence of computer navigation simultaneously on the learning curve, the frequency of intraoperative cutting errors, and component alignment in TKR. The strong points of this study include the use of a standardized surgical protocol in a single orthopedic department and the application of strict inclusion criteria. Obese patients and those with a major preoperative knee deformity were excluded. As such, it is the first study reported in the literature in which an attempt was made to reduce the influence of patient variables on the final result by minimizing these differences preoperatively. A potential weakness of the trial was that the series magnitude was not confirmed by a preliminary power study.

Using a computer navigation system reduces the influence of cutting block stability and saw blade movement on the final result. A reduction in the number of cutting errors has been shown to occur when a navigation system is used for TKR surgery [10]. In agreement with the previous study, we have shown that experience with computer navigation in TKR results in a lower number of intraoperative cutting errors. The number of re-cuts required was greater in the two groups operated on by surgeons with no prior experience in computer-assisted TKR. A statistically significant increase in the number of re-cuts was seen for TKRs performed by the general orthopedic surgeon compared with the surgeon experienced in computer-guided surgery, but we did not find any statistical difference among the group operated on by the CAS-trained surgeon and the other two groups after 11 cases.

Superior alignment and clinical results have been achieved using computer-guided TKR when compared to traditional techniques, even in experienced hands [17, 21–23]. The advantages of computer-guided TKR have not been as clearly demonstrated in low-volume surgical centers. In a retrospective study, Yau et al. [18] did not find any improvement in postoperative TKR alignment with the use of a navigation system in a low-volume practice. The authors stated that the severity of the preoperative deformity affected overall alignment postoperatively. Slover et al. [24] used a Markov decision model to demonstrate that computer navigation is less likely to be a cost-effective investment in healthcare improvement in centers with a low volume of joint replacements.

In our study, the postoperative mechanical axis of the knee was significantly better when the surgeon experienced in computer-assisted TKR performed the surgery. Other postoperative radiological parameters in the coronal plane were similar in all three groups. The accuracy of the tibial and femoral slope cut was affected by the experience of the surgeon. A statistically significantly inferior result was obtained for both of these parameters when the TKR was performed by the general orthopedic surgeon in this study. A possible explanation for this difference, based on the authors’ previous experience with computer-assisted TKR, is that saw inclination is not completely controlled by the cutting block in the sagittal plane. As a result, experience of knee replacement surgery may play an extremely important role in determining the tibial and femoral slopes in particular. Despite this, the overall postoperative TKR alignment was similar for all three surgeons. Each surgeon had a similar number of total outliers, with no statistically significant difference in the number of patients with malalignment exceeding 3° of the target value.

In 2008, Sampath et al., using a computer-assisted TKR, reported that tourniquet time increased with larger preoperative deformities and a high body mass index, and decreased with surgical experience [21]. Previous studies [12, 17, 18, 21] have shown a significant difference in surgical time, measured between skin incision and tourniquet release, when comparing inexperienced surgeons with those familiar with computer navigation. The current study also demonstrated that surgical time decreased significantly with experience in navigation, but that the intraoperative complication rate did not change.

This study shows that the learning curve needed to perform a TKR with a navigation system is 9 cases for a TKR-trained surgeon and 16 cases for a surgeon who is untrained in both CAS and TKR .


The authors demonstrated in this study that TKR performed with computer navigation yielded similar postoperative results in terms of overall alignment, even when there were variations in surgical experience. The best recovery of the mechanical axis was achieved when the surgery was performed by a surgeon experienced in computer-assisted TKR. Experience in knee replacement surgery in general leads to statistically superior tibial and femoral slopes when computer navigation is used. Experience with computer-assisted alignment techniques reduces surgical time.

In conclusion, computer navigation appears to be a useful tool in knee replacement surgery, independent of surgical experience, as surgeons with different levels of experience produced the same number of outliers. This study shows that a beginner in TKR can reproduce the correct alignment of a total knee arthroplasty just like an expert TKR surgeon (i.e., there is no statistical difference in the results achieved by the beginner and the expert surgeon) after operating on 11 cases by means of computer navigation. Initially, the surgical time is obviously longer for surgeons with little experience of CAS, but the break-even point corresponded to 16 cases. Therefore, the learning curve achieved with CAS is not as long as the traditional learning curve.

Technology now allows us to minimize human error in all areas, above all in complex systems. In our experience, even in computer-navigated joint replacement surgery, presenting precise numbers for angles, axes, and spaces can help the surgeon to standardize the surgical procedure. Furthermore, the surgeon must always perform the same surgical steps and use the same controls, like a check list. Every step is recorded, so we have an authentic black box that can show whether a bad result is a genuine human error. CAS is also a very important instructor for young surgeons; they can review every step after surgery, thus gaining an understanding of their mistakes throughout their training.

Longer follow-up will be needed to determine whether better postoperative alignment results in superior clinical outcomes and compensates for higher costs and longer surgical times.



Dr. Sunil Rajan is a well known joint replacement surgeon in Indore India. Over the last 20 years Dr. Rajan has put his expertise to work transforming more than 10,000 lines through successful knee transplants. Dr. Rajan offers his operative services at Apollo hospital Indore, one of the largest and most respected multi specialty groups in India.

Dr. Rajan offers knee replacement, hip replacement, shoulder replacement surgery as well as revision surgeries. He is one of the renowned and most experienced Doctor for knee pain in Indore India. He is passionately involved in training orthopedic surgeries in performing total and partial knee replacement surgery.  For further information visit our website or Book Appointment through online. https://www.drsunilrajan.com or call us at  9826200015.



Please go through our social media :

like our page to know more about joint replacement surgeon.


Please do follow on Instagram



Comments

Popular posts from this blog

Benefits of robotic knee replacement

Total Knee Replacement: A Panacea for Knee Problems

Benefits of Robotic knee Replacement